6GHz intelligent series radar level meter is suitable for non-contact continuous measurement of level of liquid, slurry, granule and lump material, suitable for temperature and pressure changes; occasions where inert gas and volatilization exist.
It adopts the measurement method of microwave pulse, and can work normally in the industrial frequency band range. Low beam energy, can be installed in a variety of metal, non-metallic containers or pipes, no harm to human body and the environment.
Very short microwave pulses of very low energy are transmitted and received through an antenna system. The radar wave travels at the speed of light. The running time can be converted into a level signal by means of electronic components. A special time-stretching method ensures stable and accurate measurements in a very short time. Even in the case of complex conditions and spurious echoes, the level echoes can be accurately analyzed using microprocessor technology and commissioning software.
The antenna receives the reflected microwave pulses and transmits them to the electronics. The microprocessor processes this signal and identifies the echoes generated by the micropulses on the material surface. The correct identification of the echo signal is done by the pulse software with an accuracy of millimeters.
The distance D from the material surface is proportional to the time travel T of the pulse: D=C×T/2; where C is the speed of light. Since the distance E from the empty tank is known, the level L is: L = E - D.
It is set by entering the empty tank height E (=zero), full tank height F (=full range) and some application parameters, which will automatically adapt the instrument to the measurement environment. Corresponds to 4-20mA output.
This instrument fully meets the requirements of protection class IP66/67. Please make
sure that the cable sealing head is waterproof. As shown
How to ensure that the installation meets the requirements of IP67:
Please make sure that the sealing head is not damaged.
Please make sure that the cable is not damaged.
Please ensure that the cable used meets the requirements of the electrical
connection specification.
Before entering the electrical connection, bend the cable downwards to
ensure that water does not flow into the housing, see ①
Please tighten the cable sealing head, see ②
Please plug the unused electrical connections with blind plugs, see ③
Applicable medium: liquid, slightly corrosive liquid
Applications: sewage level measurement; acid and alkali level measurement with small volatility; slurry level measurement
Explosion-proof certification: Exia IIC T6 Ga/ Exd IIC T6 Gb
Measuring range: 20m
Antenna: Rod antenna (PP/PTFE)
Frequency: 6 GHz
Process temperature: (-40 to 130)°C
Measurement accuracy: ±10mm
Process pressure: (-0.1 to 0.3) MPa
Signal output: (4 to 20)mA/HART
Field display: four-bit LCD programmable
Power supply: two-wire system (DC24V) four-wire system (DC24V/AC220V)
Repeatability: ± 1mm
Housing: aluminum single chamber/aluminum double chamber/plastic/stainless steel single chamber
Process connection: flange(optional)/thread
General data : Material Antenna PTFE, PP
Housing: Aluminum, plastic, antistatic PP, stainless steel 316
Seal between housing and enclosure: Silicone rubber
Housing window: Polycarbonate
Grounding terminal: Stainless steel
Process connection: KD-801 G1½″A thread or 1½″NPT thread/flange
KD-802 PTFE Flanged Flange
KD-803 Stainless Steel Flange
KD-804 Stainless Steel Flange
KD-805 Stainless Steel Flange
KD-806 Stainless Steel Flange
Supply voltage: Two-wire system Standard type (16~26)V DC
Intrinsically safe :(21.6~26.4)V DC
Power consumption: max. 22.5mA
Allowable ripple
- <100Hz Uss < lV
- (100~100K)Hz Uss < l0mV
Explosion-proof: (22. 8 to 26. 4) V DC 2-wire
(198~242) V AC 4-wire / 110V AC 4-wire
Power consumption : max. 1VA, 1W
Cable parameters: Cable inlet/plug 1 M20xl.5 cable inlet (cable diameter 5 to 9 mm)
1 blind plug M20xl.5
Spring terminal block: for wire cross-section 2.5 mm²
Output parameters: Output signal (4 to 20)mA/HART
Resolution : 1.6μA
Fault signal: Current output constant; 20.5mA; 22mA; 3.9mA
Two-wire load resistance: See the chart below
Four-wire load resistance: 500Ω max.
Integration time: (0~36)s, adjustable
Characteristic parameters: Blind area Antenna end
KD-801 20m (liquid)
KD-802 20m (liquid)
KD-803 35m
KD-804 35m
KD-805 20m
KD-806 15m
Measurement interval: approx. 1s (depending on parameter setting)
Adjustment time :Approx. 1s (depending on parameter setting)
Resolution :1mm
Accuracy :10mm or <0.1% (see accuracy diagram below)
Operating storage and transport temperature : (-40 to 80) ℃
Process temperature : (temperature of antenna part)
KD-801 -40~130℃
KD-802 -40~130℃ (standard type) / -40~180℃ (high temperature type)
KD-803 -40~130℃ (Standard type) / -40~180℃ (High temperature type)
KD-804 -40~130℃ (standard type) / -40~180℃ (high temperature type)
KD-805 -40~130℃ (standard type) / -40~180℃ (high temperature type)
KD-806 (-40~400)℃
Relative humidity: <95%
Pressure inside the tank: Max. 4MPa
Vibration resistance: Mechanical vibration l0m/s² , (10~150)Hz
Recommended location (2), the tank wall to the outer wall of the installation short pipe should meet the following requirements: the best distance from the tank wall is 1/4 or 1/6 of the tank diameter, the minimum installation distance from the tank wall is 1/10 of the measurement range. for example: 10m level tank, the minimum installation distance from the tank wall should be 1m.
● Cannot be installed above the inlet (4)
● It cannot be installed in the center (3). If it is installed in the center, multiple false echoes will be generated and the interference echoes will cause the real signal to be lost.
● If the distance between the instrument and the tank wall cannot be maintained, the medium on the tank wall will adhere to cause false echoes, and false echo storage should be performed when commissioning the instrument.
Our main products are: high frequency radar level meter, radar level meter, guided wave radar level meter and other instrumentation products, if you need, please contact us!
We are here to help you! If you close the chatbox, you will automatically receive a response from us via email. Please be sure to leave your contact details so that we can better assist